Instrument Functions
R&S
®
NGP800
64User Manual 5601.5610.02 ─ 04
Figure 6-13: Arbitrary editor dialog
1. Configure the "Arb Editor" with the required voltage, current and duration.
The R&S NGP800 displays the on-screen keypad for data entry.
2. Confirm values with the unit keys.
3. Select the interpolation checkbox to toggle on/off the interpolation function on the
arbitrary data.
4. Select the "Plus" or "Minus" icon to add or delete the arbitrary data from the dialog.
"+": A new row of arbitrary data is added to the end of the table. The new addi-
tion is a copy of the last arbitrary data in the table.
"-": To delete, select the row of arbitrary data for deletion follows by the "Minus"
icon.
5. Set the "Rep" to configure repetition cycle for the arbitrary data.
By default, the repetition cycle is set to infinity.
6. Set the "End Behavior" to handle the way to end the automation of the QuickArb
function.
"Off": The output of the selected channel is turned off after performing the
QuickArb function.
"Hold": The last voltage and current values remains at the output terminal of
the instrument.
7.
Select
(new file) or (existing file) to save the arbitrary data.
The R&S NGP800 displays a popup message to show that file saved successfully.
Advanced Features