Remote Control Commands
R&S
®
NGP800
157User Manual 5601.5610.02 ─ 04
CALibration:AINPut:UMIN
Sets the output voltage to low value 1 % of Vmax for analog input pin during adjust-
ment.
Usage: Event
CALibration:CANCel
Cancels the channel adjustment.
Usage: Event
CALibration:COUNt?
Queries the number of counts channel adjustment performed successfully.
Usage: Query only
CALibration:CURRent:DATA <arg0>
Set the DMM reading after setting the output current level in channel adjustment proc-
ess.
Parameters:
<current> Measured value from DMM.
CALibration:CURRent:IMAX
Sets the output current to high value 100 % of Imax during current adjustment.
Usage: Event
CALibration:CURRent:IMIN
Sets the output current to low value 1 % of Imax during current adjustment.
Usage: Event
CALibration:DATE?
Returns the channel adjustment date.
Usage: Query only
CALibration:END
Ends the channel adjustment.
Usage: Event