Rohde & Schwarz NGP814 power supply

NGP800 series manual - Page 109

For NGP814.

PDF File Manual, 190 pages, Read Online | Download pdf file

NGP814 photo
Loading ...
Loading ...
Loading ...
Remote Control Commands
R&S
®
NGP800
107User Manual 5601.5610.02 ─ 04
OTP
Output the selected logic level when the selected critical event
(OTP) occurs on the selected channel.
OPP
Output the selected logic level when the selected critical event
(OPP) occurs on the selected channel.
VMODe
Output the selected logic level when the selected channel oper-
ates in the CV mode.
CMODe
Output the selected logic level when the selected channel oper-
ates in the CC mode.
VLEVel
Output the selected logic level when the voltage level of the
selected channel is greater or equal to the set voltage level, i.e.
Vset >= set value.
ILEVel
Output the selected logic level when the current level of the
selected channel is greater or equal to the set current level, i.e.
Iset >= set value.
ENABle
Selected channel output is turned on when the selected logic
level is met.
INHibit
Selected channel output is inhibited when the selected logic
level is met.
Note 1: If the selected channel output is put to inhibit state, man-
ual or remote operation on selected channel output is no longer
possible .
Note 2: To reverse the inhibit state, remove the source of the
trigger signal. You can either disable the affected DIO interface
or remove the source from the affected DIO interface at the rear
panel.
ARB
Selected channel QuickArb function is enabled when the
selected logic level is met.
ARBPoint
Selected channel QuickArb function is stepped to the next point
when the selected logic level is met.
ARBGroup
Selected channel QuickArb function is stepped to the next group
when the selected logic level is met.
RAMP
Selected channel ramp is enabled when the selected logic level
is met.
Trigger Commands
Loading ...
Loading ...
Loading ...