Remote Control Commands
R&S
®
NGP800
125User Manual 5601.5610.02 ─ 04
OUTPut[:STATe] <arg0>[, <Channel list>]
OUTPut[:STATe]? [<Channel list>]
Sets or queries the output state of the previous selected channels.
Parameters:
<state> 0
Switches off previous selected channels.
1
Switches on previous selected channels.
Parameters for setting and query:
<Channel list> <list>
Example:
OUTP? (@1)
Queries the output state at channel 1.
Example: See
Example "Activating the channels" on page 124.
OUTPut:DELay:DURation <New value for sequence delay (selected channel)>[,
<Channel list>]
OUTPut:DELay:DURation? [<Channel list>]
Sets or queries the duration for output delay.
Setting parameters:
<duration> <numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Numeric value of the duration in seconds.
MIN | MINimum
Minimum value of the duration at 0.001 seconds.
MAX | MAXimum
Maximum value of the duration at 10.00 seconds.
Range: 0.001 to 10.00
*RST: 0.001
Default unit: s
Parameters for setting and query:
<Channel list> <list>
Example:
OUTPut:DELay:DURation 1
OUTPut:DELay:DURation? -> 1
Returns output delay of 1 s.
Example:
OUTPut:DELay:DURation? (@1)
Returns output delay at channel 1.