Remote Control Commands
R&S
®
NGP800
148User Manual 5601.5610.02 ─ 04
0
QuickArb function is deactivated.
*RST: 0
Parameters for setting and query:
<Channel list> <list>
Example:
ARB ON
ARB? -> 1
QuickArb function of Ch1 is activated.
See
Example "Configuring an arbitrary sequence" on page 145.
Example:
ARB ON, (@1)
ARB? (@1)
Sets and queries the state of QuickArb function at channel 1.
ARBitrary:CLEar
Clears the previous defined arbitrary waveform data for the selected channel.
Example: See
Example "Configuring an arbitrary sequence" on page 145.
Usage: Event
ARBitrary:DATA <>...
Sets or queries the arbitrary points for the previous selected channel. Max. 1024 arbi-
trary points can be defined. The dwell time between 2 arbitrary points is specified from
1 ms to 60 ms.
Parameters:
<data> voltage1, current1, time1, interpolation mode1, voltage2, cur-
rent2, time2, interpolation mode2, ...
Voltage and current settings depending on the instrument type.
If the interpolation mode is sets to 1, it indicates that the mode is
activated. If the interpolation mode is sets to 0, it indicates that
the mode is not activated.
Example:
INST OUT1
ARB:DATA 10,1,0.5,0
Defines one arbitrary point with: Voltage1 = 10 V and Current1 =
1 A, Time1 = 500 ms and Interpolation mode1 = 0 (disabled).
ARB:DATA? -> 10.000, 1.000, 0.50, 1
Returns defined arbitrary points for the previous selected chan-
nel.
See
Example "Configuring an arbitrary sequence" on page 145.
ARBitrary:SEQuence:ENDPoint?
Queries the total number of points of the arbitrary sequence.
Usage: Query only
Advanced Operating Commands