Rohde & Schwarz NGP814 power supply

NGP800 series manual - Page 27

For NGP814.

PDF File Manual, 190 pages, Read Online | Download pdf file

NGP814 photo
Loading ...
Loading ...
Loading ...
Getting Started
R&S
®
NGP800
25User Manual 5601.5610.02 ─ 04
9 = AC inlet with integrated 2-pole rocker switch
10 = Optional IEEE-488 (GPIB) interface
11 = Ground terminal
12 = Kensington lock
13 = Channel 3 and 4 rear panel connector (for NGP804, NGP814 and NGP824 models only)
14 = Channel 1 and 2 rear panel connector
15 = Analog input and digital I/O connector
16 = USB-B connector (device)
17 = Ethernet (LAN) connector
AC inlet with integrated 2-pole rocker switch (9)
Main supply cable
Use only the power cable that was supplied with the instrument. Using other types,
which might have inadequate rating can cause overheating of the power cable, result-
ing in fire.
The power cable provides the earth ground connection through the third ground con-
ductor. Operate the instrument only on authorized safety sockets which provide earth
connection.
The power cable must be plugged in before signal circuits can be connected. Never
use the product if the power cable is damaged.
The built-in 2-pole rocker switch is the main power switch of the instrument which con-
nects/disconnects it from the AC supply.
Option IEEE-488 (GPIB) interface (10)
Option R&S NG-B105 provides an IEEE-488 (GPIB) bus interface.
Ground terminal (11)
M4 screw provides connection to earth ground through the instrument ground/chassis.
Kensington security slot (12)
A Kensington lock can be anchored to the R&S NGP800 power supply housing to
secure it to a workstation mechanically.
Channel connectors (13, 14)
Output terminals
Either the channel output terminals at the front panel or rear panel can be used. Using
both terminals at the same time can cause instrument malfunction.
Instrument Tour
Loading ...
Loading ...
Loading ...