Rohde & Schwarz NGP814 power supply

NGP800 series manual - Page 138

For NGP814.

PDF File Manual, 190 pages, Read Online | Download pdf file

NGP814 photo
Loading ...
Loading ...
Loading ...
Remote Control Commands
R&S
®
NGP800
136User Manual 5601.5610.02 ─ 04
0
OPP is deactivated
1
OPP is activated
Parameters for setting and query:
<Channel list> <list>
Example:
POW:PROT? (@1)
Queries OPP state at channel 1.
Example: See
Example "Configuring the overpower protection"
on page 135.
[SOURce:]POWer:PROTection:CLEar [<Channel list>]
[SOURce:]POWer:PROTection:CLEar [<Channel list>]
Resets the OPP state of the selected channel. If an OPP event has occurred before,
the reset also erases the message on the display.
Parameters:
<Channel list> <list>
Example:
POW:PROT:CLE (@1)
Resets OPP state at channel 1.
Example: See
Example "Configuring the overpower protection"
on page 135.
Usage: Setting only
[SOURce:]POWer:PROTection:LEVel [<New value for voltage protection>, <Channel
list>]
[SOURce:]POWer:PROTection:LEVel? [<Channel list>]
Sets or queries the overvoltage protection value of the selected channel.
Parameters:
<power> <numeric value> | MIN | MINimum | MAX | MAXimum | DEF |
DEFault
<numeric value>
Numeric value of the power protection level in watts.
MIN | MINimum
Minimum value of the power protection level at 0.00 W.
MAX | MAXimum
Maximum value of the power protection level at 200.00 W.
DEF | DEFault
Default value of the power protection level at 200.00 W.
Range: 0.00 to 200.00
*RST: 200.00
Default unit: W
Configuration Commands
Loading ...
Loading ...
Loading ...