Instrument Functions
R&S
®
NGP800
63User Manual 5601.5610.02 ─ 04
7.
Select delete and up/down button
to navigate the arbitrary test sequence.
8. Select "Load Sequence" to load the arbitrary test sequence.
9. Set "Seq. Rep." and "End Behavior" to configure the arbitrary sequence behavior.
"Seq. Rep.": Repetition cycle for the arbitrary sequence
"End Behavior": End behavior of the automation of arbitrary function
"Off": Output of the selected channel is turned off after performing the
QuickArb function.
"Hold": Last voltage and current values output at the instrument.
10. Select [Back] key to return to channel menu dialog.
11. Activate the "Arbitrary" menu item to enable the QuickArb function.
The R&S NGP800 enables the QuickArb function and displays the "Arbitrary" icon
on the selected channel status bar information.
Arbitrary editor
The "Arb Editor" dialog allows you to edit the arbitrary profile ("Voltage", "Current",
"Time", "Interpolate" status, "Rep." and "End Behavior"). To view or open the list of
available arbitrary files, select
to open the arbitrary file.