Remote Control Commands
R&S
®
NGP800
153User Manual 5601.5610.02 ─ 04
MIN | MINimum
Minimum duration of the ramp function at 0.00 s.
MAX | MAXimum
Maximum duration of the ramp function at 60.00 s.
DEF | DEFault
Default duration of the ramp function at 0.01 s.
Range: 0.01 to 60.00
*RST: 0.01
Default unit: s
Parameters for setting and query:
<Channel list> <list>
Example:
VOLT:RAMP:DUR 4
VOLT:RAMP:DUR? -> 4
Duration of the ramp function is set at 4 s.
Example:
VOLT:RAMP:DUR 4, (@1)
VOLT:RAMP:DUR? (@1)
Sets and queries the duration of ramp function at channel 1.
7.7.3 Analog Input
The VOTage:AINPut subsystem contains the commands for configuring the analog
input.
[SOURce:]VOLTage:AINPut:INPut................................................................................... 153
[SOURce:]VOLTage:AINPut:TRIGgered[:STATe ]...............................................................154
[SOURce:]VOLTage:AINPut[:STATe]................................................................................ 154
[SOURce:]VOLTage:AINPut:INPut <arg0>[, <Channel list>]
[SOURce:]VOLTage:AINPut:INPut? [<Channel list>]
Sets or queries the analog input mode.
Parameters:
<input> VOLT | CURR | <list>
VOLT
Voltage mode.
CURR
Current mode.
Parameters for setting and query:
<Channel list> <list>
Example:
VOLT:AINP:INP? (@1)
Queries the analog input mode at channel 1.