Remote Control Commands
R&S
®
NGP800
156User Manual 5601.5610.02 ─ 04
CALibration:AINPut:SAVE
Saves the analog input adjustment.
Usage: Event
CALibration:AINPut:STARt <arg0>
Selects the analog input pin for adjustment.
Setting parameters:
<pin> Input pin for adjustment.
Range: 1 to 4
CALibration:AINPut:STATe?
Queries the analog input adjustment state.
State Descriptions
0-15 0x0 - 0xE ( 0x0000 - 0x1111)
bit3|bit2|bit1|bit0
bit0 - pin 1 of analog input
bit1 - pin 2 of analog input
bit2 - pin 3 of analog input
bit3 - pin 4 of analog input
e.g. 15 - All analog input pins are adjusted.
e.g. 9 - Pin 1 and pin 4 are adjusted.
16 Idle
17 Busy
18 Waiting
Example:
CAL:AINP:STAT? -> 9
Pin 1 and pin 4 are adjusted successful.
Usage: Query only
CALibration:AINPut:UMAX
Sets output voltage to high value 100 % of Vmax for analog input pin during adjust-
ment.
Usage: Event