Instrument Functions
R&S
®
NGP800
52User Manual 5601.5610.02 ─ 04
AUTO : The detection and enabling of the voltage sense relay automatically
kicks in when the connection of remote sensing wires (S+, S- ) to the input of
the load is applied. The voltage sensing relay is switched off when output is
turned off in this case.
3. Select "Set" to configure the remote sensing mode.
6.3 Analog Input
Instrument option
R&S NGP-K107 (P/N: 5601.6200.03) option is required for the "Analog Input".
Analog input
If "Analog Input" of a selected channel (voltage or current) is enabled, the respective
channel voltage or current setting is disabled.
See
Chapter 6.1, "Setting the Channels Voltage and Current", on page 47.
With "Analog Input", you can regulate the channel output (voltage or current) settings
with an input voltage of 0 V to 5 V.
Analog input mode NGP output for 64 V channels NGP output for 32 V chan-
nels
Voltage mode 0 V to 64 V 0 V to 32 V
Current mode 0 A to 10 A 0 A to 20 A
1. Press [Settings] key.
The R&S NGP800 displays the device/channel menu window.
2. Select the required channel tab to configure the analog input mode.
The R&S NGP800 displays the selected channel menu.
3. Select "Analog Input" from the menu.
The R&S NGP800 displays the selected "Analog Input" dialog.
Analog Input