Instrument Functions
R&S
®
NGL200/NGM200
68User Manual 1178.8736.02 ─ 09
4. Activates the "Enabled" menu item.
The R&S NGL/NGM enables the Arbitrary function and displays the "Arbitrary" icon
on the selected channel status bar information.
5. Select "Load from file.." to load the arbitrary file.
6. Select the required source and file location.
Alternatively, select "New file" to edit a new arbitrary file. The R&S NGL/NGM
opens the
arbitrary editor dialog to edit the arbitrary file.
7. Select "Load" to load the selected file.
The R&S NGL/NGM loads the selected arbitrary file.
8. Applicable only with R&S NGL-K103 or R&S NGM-K103, Digital Output option.
See Chapter 6.6, "Trigger / Digital I/O", on page 62.
Activate "Start triggered" or "Step triggered" if the arbitrary function is executed
under triggered condition.
"Start triggered": If triggered, the complete arbitrary function is executed.
"Step triggered": If triggered, every arbitrary step needs a trigger signal to exe-
cute (step time from "Arb Editor" is ignored).
Arbitrary editor
Interpolation flag
After execution of all repetition cycles, the last interpolation flag is ignored. See
Fig-
ure 6-22.
The "Arb Editor" dialog allows you to edit the arbitrary profile ("Voltage", "Current",
"Time", "Interpolate" status, "Rep." and "End Behavior"). To view or open the list of
available arbitrary files, select
to open the arbitrary file.