Instrument Functions
R&S
®
NGL200/NGM200
46User Manual 1178.8736.02 ─ 09
Figure 6-3: Output of Ch2 in CV mode
6.2.1 Set Constant Resistance
By enabling the constant resistance (CR) mode, you can operate the R&S NGL/NGM
as an electronic load in sink mode. This allows you to perform testing that requires a
constant load resistor in your application.
1. Press [Settings] key.
The R&S NGL/NGM displays the device/channel menu window.
2. Select the required channel tab to apply constant resistance.
The R&S NGL/NGM displays the selected channel menu.
3. Select the "Constant Resistance" from the menu.
The R&S NGL/NGM displays the "Constant Resistance" dialog.
4. Activate the "Enabled" menu item.
When operates in sink mode condition, the R&S NGL/NGM displays the "Constant
Resistance" icon on the selected channel status bar information.
5. Set the required resistance.
The R&S NGL/NGM displays the on-screen keypad to set the value.
6. Confirm value with the unit key (mΩ or Ω).
The value for "Constant Resistance" is rounded to the nearest 1 Ω or 0.1 Ω,
depending on the value.
Activating the Channels Output