Rohde & Schwarz NGM-COM2B Rohde &Schwarz power supply bundle

User manual NGM COM2a - Page 121

For NGM-COM2B.

PDF File Manual, 201 pages, Read Online | Download pdf file

NGM-COM2B photo
Loading ...
Loading ...
Loading ...
Remote Control Commands
R&S
®
NGL200/NGM200
119User Manual 1178.8736.02 ─ 09
DOWN
Decreases resistance by a defined step size.
Increment: 0.1 ohms
Default unit: ohms
Parameters for setting and query:
<Channel list> <list>
Example:
RES? (@1)
Queries the constant resistance target value at channel 1.
Example: See
Example "Configuring the resistance limit" on page 118.
[SOURce:]RESistance:STATe <state>
[SOURce:]RESistance:STATe? [<Channel list>]
Sets or queries the constant resistance mode.
Parameters:
<state> 0
Deactivates constant resistance mode.
1
Activates constant resistance mode.
Parameters for setting and query:
<Channel list> <list>
Example:
RES:STAT? (@1)
Queries the constant resistance mode at channel 1.
Example: See
Example "Configuring the resistance limit" on page 118
7.5.6 Combined Setting of Voltage and Current Setting
The APPLy subsystem provides a command that enables you to set the current and
voltage of a channel in one step.
The combined voltage and current setting command takes approximately 100 ms, i.e.
longer than the setting of a single value.
APPLy <voltage> [,<current>][,<output>]
Sets or queries the voltage and current value of the selected channel.
Parameters:
<voltage> <numeric value> | MIN | MINimum | MAX | MAXimum | DEF |
DEFault
<numeric value>
Numeric value for voltage in the range of 0.000 to 20.050.
MIN | MINimum
Min voltage at 0.000 V.
Configuration Commands
Loading ...
Loading ...
Loading ...