Rohde & Schwarz NGM-COM2B Rohde &Schwarz power supply bundle

User manual NGM COM2a - Page 122

For NGM-COM2B.

PDF File Manual, 201 pages, Read Online | Download pdf file

NGM-COM2B photo
Loading ...
Loading ...
Loading ...
Remote Control Commands
R&S
®
NGL200/NGM200
120User Manual 1178.8736.02 ─ 09
MAX | MAXimum
Max value for voltage at 20.050V.
DEF | DEFault
Default voltage.
*RST: 1.000
Default unit: V
<current> <numeric value> | MIN | MINimum | MAX | MAXimum | DEFault
<numeric value>
Numeric value for current in the range of 0.000 to 6.010.
MIN | MINimum
Min current at 0.000 A.
MAX | MAXimum
Max value for current at 6.010 A.
DEF | DEFault
Numeric value for current.
*RST: 1.000
Default unit: A
<output> OUT1 | OUTP1 | OUTPut1 | OUT2 | OUTP2 | OUTPut2
OUT1 | OUTP1 | OUTPut1
Selects output for channel 1.
OUT2 | OUTP2 | OUTPut2
Selects output for channel 2.
Example:
INST OUT1
APPL 6,2
Sets 6 V and 2 A to output of channel 1.
APPL? -> 6.000, 2.000
Queries the voltage and current of the selected channel.
READ? [<Channel list>]
Queries the currently voltage-current measurement pair and returns them separated by
comma.
Parameters:
<Channel list> <list>
Example: READ? -> 1.001080E+00,1.004224E-02
Example:
READ? (@1)
Queries the voltage-current measurement pair at channel 1.
Usage: Query only
7.5.7 Output Setting
The OUTPut subsystem contains the commands for activating the output channels.
Configuration Commands
Loading ...
Loading ...