Instrument Functions
R&S
®
NGL200/NGM200
51User Manual 1178.8736.02 ─ 09
The R&S NGL/NGM displays the "Trigger" dialog.
3. Select the required trigger action.
"Output On": If triggered, the selected output channel is turned on.
"Output Off": If triggered, the selected output channel is turned off.
"Gated": If triggered, the selected output channel is gated.
4. Select "Set" to configure the trigger action.
6.2.3.4 Output Mode
1. Select the "Output Mode" menu item to set the output mode. See "Source and sink
current" on page 26.
The R&S NGL/NGM displays the "Output mode" dialog.
2. Select the required output mode.
Auto: Depending on the voltage across the output terminal, the R&S NGL/NGM
goes into sink or source mode.
Sink : The R&S NGL/NGM goes to sink mode, current flows into the instru-
ment. On display, the current shown negative current.
Source: The R&S NGL/NGM goes to source mode, current flows out from the
instrument.
Note: If "Sink" mode is selected, the source current is limited to 1 mA. Otherwise, if
"Source" is selected, the sink current is limited to 1 mA.
3. Select "Set" to configure the trigger condition.
Activating the Channels Output