Remote Control Commands
R&S
®
NGL200/NGM200
170User Manual 1178.8736.02 ─ 09
#<LengthofLength><Length><block_data>
Example: #234<block_data>
<LengthofLength> specifies how many positions the subsequent length specifica-
tion occupies ("2" in the example)
<Length> specifies the number of subsequent bytes ("34" in the example)
<binary block data> specifies the binary block data of the specified length
To configure fastlog for scpi target, see
Example "Configuring fastlog for scpi target"
on page 167.
Example:
FLOG:DATA?
Returns the binary format of fastLog data from channel 1.
Usage: Query only
FLOG:SRATe <samplerate>
Sets or queries the sample rate of the FastLog function.
Parameters:
<samplerate> S500k | S250k | S50k | S10k | S1k | S100
Example:
FLOG:SRAT S500K
Set Ch1 FastLog writes sample rate at 500K samples/sec.
FLOG:TRIGgered <arg0>
FLOG:TRIGgered?
Sets or queries the triggered state of FastLog.
See Figure 6-18.
Parameters:
<arg0> 1
Activates the FastLog state.
0
Deactivates the FastLog state.
Example:
FLOG:TRIG 1
Ch1 FastLog triggered state is activated. In the event if a trigger
signal is detected from the various input digital sources, a trigger
event is sent to activate the FastLog function.
FLOG:FILE:DURation <arg0>
FLOG:FILE:DURation?
Sets or queries the file write duration.
Setting parameters:
<duration> Sets file write duration.