Remote Control Commands
R&S
®
NGL200/NGM200
132User Manual 1178.8736.02 ─ 09
*RST: 0
Default unit: s
Parameters for setting and query:
<Channel list> <list>
Example:
CURR:PROT:DEL:INIT? (@1)
Queries initial fuse delay time at channel 1.
Example: See
Example "Configuring the overcurrent protection"
on page 129.
[SOURce:]CURRent:PROTection:DELay <New value for voltage>[, <Channel list>]
[SOURce:]CURRent:PROTection:DELay? [<Channel list>]
Sets or queries the fuse delay time.
Parameters:
<duration> <numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Numeric value for the initial fuse delay.
MIN | MINimum
Min value for initial fuse delay.
MAX | MAXimum
Max value for initial fuse delay.
Range: 0.00 to 10.00
*RST: 0
Default unit: s
Parameters for setting and query:
<Channel list> <list>
Example:
CURR:PROT:DLEAY 1, (@1)
Sets initial fuse delay 1 s at channel 1.
Example: See
Example "Configuring the overcurrent protection"
on page 129.
[SOURce:]CURRent:PROTection:LINK <arg0>
[SOURce:]CURRent:PROTection:LINK? <arg0>
Sets or queries the fuses of several selected channels (fuse linking).
Parameters for setting and query:
<arg0> 1 | 2
Example:
INST OUT1;:CURRent:PROTection:LINK 2
Channel 2 is linked with channel 1
INST OUT1;:CURRent:PROTection:LINK?2
Returns 1 if channel 1 is linked with channel 2
Example: See
Example "Configuring the overcurrent protection"
on page 129.
Configuration Commands