Instrument Functions
R&S
®
NGL200/NGM200
48User Manual 1178.8736.02 ─ 09
3. Activate the "Fast Transient Response" from the menu.
The R&S NGL/NGM applies the fast transient response on the operating condition
and displays the "Fast Transient Response" icon on the selected channel status
bar information.
The setting of "Fast Transient Response" affects the output voltage control loop
behavior. See
Chapter 6.2.3.1, "Impedance", on page 48.
If activated, the focus is on control loop speed, which is recommended for resistive
loads. Otherwise, the focus is on stability and should be preferred for capacitive or
inductive loads.
6.2.3 Output
The "Output" menu provides the settings for output impedance, output delay and trig-
gers action
on the output mode.
1. Press [Settings] key.
The R&S NGL/NGM displays the device/channel menu window.
2. Select the required channel tab to apply output delay.
The R&S NGL/NGM displays the selected channel menu.
3. Select the "Output" menu item.
The R&S NGL/NGM displays the "Output" dialog.
Figure 6-6: Output delay dialog
6.2.3.1 Impedance
The output impedance function is disabled during voltage ramp time of "Ramp" func-
tion. See Chapter 6.7.2, "Ramp", on page 70.
1. Select the "Impedance" menu item to configure the required values.
Activating the Channels Output