SSA3000X Plus user manual - Page 99

For SSA3015X-PLUS.

Loading ...
Loading ...
Loading ...
SIGLENT
User Manual 77
Figure 3-6 Calibration Wizard
3.5.6.1 Correction
Turn on or off the calibration.
3.5.6.2 Calibrate
1. 1-Port Cal
Calibrate with the specified mechanical calibration kit, and need to connect an OPEN standard, a
SHORT standard, and a LOAD standard to the [Port 1]. This calibration effectively eliminates the
frequency response reflection tracking error, directivity error, and source match error from the test setup
in a reflection test using that port.
This function is only available when the measurement item is S11.
Calibration data is saved as user calibration data.
2. Response Through
When operating, connect the port 1 and port 2 of the analyzer with an optional Through Adapter. The
normalization operation moves the measurement reference plane to both ends of the Through Adapter.
This function is only available when the measurement item is S21. Enter the normalized submenu to
make the appropriate selection.
3.5.6.3 Ecal
Use the optional SIGLENT electronic calibration kit for calibration. Calibration data is saved as user
calibration data.
3.5.6.4 Cal Kit
Specify the calibration kit used for mechanical calibration.
F503ME:Type-N 50Ω 3 GHz Cal Kit (Siglent)
85032F:Type-N 50Ω 9 GHz Cal Kit (KeySight)
85032B/E
User
Before executing calibration, you need to select a calibration kit.
If you use a calibration kit other than a predefined one, you need to define it. If the connector type of the
standard of the calibration kit you use has polarity (the distinction between male and female), you need