SSA3000X Plus user manual - Page 147

For SSA3015X-PLUS.

Loading ...
Loading ...
Loading ...
SIGLENT
User Manual 125
Table 7-18 Marker parameters
Parameter
Explanation
Default
Center Frequency
Range
0 ~ Full Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Step = Span/(Sweep Points - 1)
Direction Key Step
Step = Span/10
7.4.1.2 Select Trace
Select the trace to be marked by the current marker. Valid selections include A, B, or C.
7.4.1.3 Normal
One of the marker types. It is used to measure the X (Frequency or Time) and Y (Amplitude) values of
a certain point on the trace. When selected, a marker with the number of the current marker (such as
1”) appears on the trace.
If no active marker exists currently, a marker will be enabled automatically at the center frequency
of the current trace.
You can use the numeric keys, knob or direction keys to move the marker. The readouts of the
marker will be displayed at the upper right corner of the screen.
The readout resolution of the X-axis (frequency or time) is related to the span. For higher readout
resolution, reduce the span.
7.4.1.4 Delta
One of the marker types. It is used to measure the delta values of X (Frequency or Time) and Y
(Amplitude) between the reference point and a certain point on the trace. When selected, a pair of
markers appears on the trace: Fixed Related Marker (marked by a combination of the marker number
and letter “+”, such as “2+”) and the Delta Marker (marked by the “∆”, such as “1∆2”).
After the marker selects “Delta”, the original marker will become the delta measurement marker,
and the related marker of the incrementing sequence number will become the reference “fixed”
marker
The delta marker is in the "relative to" state, and its X-axis position can be changed; the related
marker is in the "fixed" state by default (the X-axis and Y-axis positions are fixed), but the X-axis
can be adjusted by changing to the "normal" state.
The first row in the upper right corner of the trace area shows the frequency (or time) difference
and amplitude difference between the two markers; the second row in the upper right corner of the
trace area shows the X axis and amplitude value of the related marker.
7.4.1.5 Fixed
One of the marker types. When “Fixed” is selected, the X-axis and Y-axis of the marker will not change
Loading ...
Loading ...
Loading ...