154 User Manual
Figure 9-2 LAN
4. Press button on the front panel System→Interface→LAN to enter the LAN Config function menu.
5. Select the IP Config between Static and DHCP.
DHCP: the DHCP server in the current network will assign the network parameters automatically
(IP address, subnet mask, gate way) for the analyzer.
Static: you can set the IP address, subnet mask, gate way manually. Press Apply.
Figure 9-3 LAN Config
The analyzer will be detected automatically or manually as a new LAN point.
9.1.3 GPIB: Connecting the Analyzer via the USB-Host port
Refer to the following steps to finish the connection via USB:
1. Install NI-VISA on your PC for GPIB driver.
2. Connect the analyzer USB Host port to a PC’s GPIB card port, with SIGLENT USB-GPIB adaptor.
3. Switch on the analyzer
Figure 9-4 SIGLENT USB-GPIB Adaptor
4. Press button on the front panel System→Interface→GPIB to enter the GPIB number.
The analyzer will be detected automatically as a new GPIB point.