SSA3000X Plus user manual - Page 136

For SSA3015X-PLUS.

Loading ...
Loading ...
Loading ...
SIGLENT
114 User Manual
7.2 Basic Settings
7.2.1 Frequency
Set the frequency-related parameters and functions of the analyzer. Any change of this parameter will
stop the running sequence.
7.2.1.1 Freq(Meter)
Set the frequency of the meter measurement. The meter frequency value is displayed at the bottom of
the meter window.
The frequency of the meters can be at any position, even outside the span displayed on the screen.
Table 7-1 Meter Frequency
Parameter
Explanation
Default
165MHz
Range
0 Hz ~ Full Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Span/200, min 1 Hz
Direction Key Step
Span/10
Related to
7.2.1.2 Midspan Frequency
Set the frequency at the midspan of scan.
Modifying the center frequency will modify both the start frequency and stop frequency when the
span is constant (except when the start frequency or stop frequency reaches the boundary).
Table 7-2 Center Frequency
Parameter
Explanation
Default
165 MHz
Range
50 Hz ~ (Full Span -50Hz)
Unit
GHz, MHz, kHz, Hz
Knob Step
Span/200, min 1 Hz
Direction Key Step
Span/10
Related to
Start Freq, Stop Freq
7.2.1.3 Start Frequency
Set the start frequency of scan. The start and stop frequencies are displayed at the bottom of the grid