SSA3000X Plus user manual - Page 82

For SSA3015X-PLUS.

Loading ...
Loading ...
Loading ...
SIGLENT
60 User Manual
Table 2-42 Noise BW
Parameter
Explanation
Default
3 MHz
Range
100 Hz ~ (2*span - 2*|Freq Offset| - Carrier BW)
Unit
GHz, MHz, kHz, Hz
Knob Step
Noise BW /10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
3. Freq Offset
Set the difference between carrier center frequency and noise center frequency.
Table 2-43 Freq Offset
Parameter
Explanation
Default
3 MHz
Range
-(span - (Carrier BW + Noise BW) /2) ~
(span - (Carrier BW + Noise BW) /2)
Unit
GHz, MHz, kHz, Hz
Knob Step
Freq Offset /10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
2.4.2.9 Harmonics
Figure 2-22 Harmonics
Measurement Results: each harmonic amplitude and total harmonic distortion of carrier signal. It can
measure up to 10th harmonic.
Measurement Parameter: