SSA3000X Plus user manual - Page 122

For SSA3015X-PLUS.

Loading ...
Loading ...
Loading ...
SIGLENT
100 User Manual
analyzer always captures all the data within a specific time interval and processes (Peak, Average, etc.)
the data using the detector currently selected, then it displays the processed data (a single data point)
on the screen.
Select an appropriate detector type according to the actual application in order to ensure the accuracy
of the measurement.
The available types are Pos Peak, Neg Peak, Sample and Average. The default is Pos peak.
1. Positive Peak
For each trace point, Positive Peak detector displays the maximum value of data sampled within the
corresponding time interval.
2. Negative Peak
For each trace point, Negative Peak detector displays the minimum value of data sampled within the
corresponding time interval.
3. Sample
For each trace point, Sample detector displays the transient level corresponding to the central time point
of the corresponding time interval. This detector type is applicable to noise or noise-like signal.
4. Average
For each trace point, Average detector displays the average value of data sampled within the
corresponding time interval.
6.2.4 Sweep
Sets parameters about the Sweep functions, including acquisition time, sweep mode, sweep times, etc.
6.2.4.1 Acquisition Time
Set the acquisition time of real-time spectrum analyzer within the real-time analysis span. The
acquisition time can be set in “Auto” or “Manual” mode and the default is “Auto”.
Table 6-12 Acquisition Time
Parameter
Explanation
Default
N/A
Range
98.75us ~ 40s
Unit
ks, s, ms, us
Knob Step
Sweep time/100, min =1 ms
Direction Key Step
Increasing multiples
.