User Manual 83
4.2.5 Cable Atten
Set the attenuation factor of the cable-under-test. It is used to compensate the amplitude of peaks in
different positions. The DTF calculates the peaks by the final receiving data which has been attenuated
by the cable, thus the amplitude of peaks cannot show exactly where the mismatch position is. So the
cable atten is used to compensate by length.
Table 4-5 Cable Atten
Parameter
Explanation
Default
0.00 dB/m
Range
0.00 dB/m ~ 5.00 dB/m
Unit
dB/m
4.2.6 Window
Set the window function used in DTF.
The use of a non-rectangular window function can improve the side lobe effect of the analysis. The
vertical axis is more accurate, but the horizontal axis resolution is reduced.
In the sub-menu, the following three settings can be selected.
Off
Rectangular
Hamming
The default state is "Off"; this value is also displayed in the status bar on the left side of the screen.
Table 4-6 Window function properties
Win-type
Expression
Main lobe width
Rectangular
Hamming
4.2.7 Calibration
Set calibration related items. The calibration status is displayed in the upper left corner of the screen.
The calibration status and display are as follows:
No calibration data --- (displayed in gray)
Calibrated Cor (displayed in blue)
Need to re-calibrate C? (displayed in blue)
Note: The calibration status is shown as “C?” which indicates that the scan frequency range is different
from the time of calibration. Users need to re-calibrate in the current scan frequency range.
1. Correction