User Manual 133
8.1.3 Interface
The analyzer supports communications through LAN, USB and the SIGLENT USB-GPIB adapter as
standard remote computer control interfaces.
8.1.3.1 LAN
Configure the related parameters of the LAN connection by “Static” or “DHCP” method. As a default, the
IP config is DHCP.
Statically setting LAN parameters requires pressing the "Apply" button to make LAN parameters
effective.
When “DHCP” is pressed, the LAN parameters are dynamically allocated.
The LAN status icon will be displayed in the upper right corner of the screen if the cable is connected.
Figure 8-1 LAN Config
8.1.3.2 GPIB
Configure GPIB port number. The analyzer provides a digital interface for use with an optional
SIGLENT USB-GPIB module through the front USB port.
8.1.3.3 Web Server
Set web page parameters to use VNC.
1. Port
The port number can be set to 5900~5999. When the port number is set beyond the range, a prompt
pops up at the bottom of the screen, and the port number is automatically set to the maximum or
minimum value.
After modifying the port number, VNC needs to be reopened.