SSA3000X Plus user manual - Page 173

For SSA3015X-PLUS.

Loading ...
Loading ...
Loading ...
SIGLENT
User Manual 151
Det.1 Dwell Time
5 ms
Det.1 Limit for
1
Det.2 Switch
On
Det.2 Dwell Time
5 ms
Det.2 Limit for
1
Det.3 Switch
On
Det.3 Dwell Time
5 ms
Det.3 Limit for
1
Meter Dwell Time
10 ms
Meter 1
On, Peak
Meter 2
On, QPeak
Meter 3
On, EMI Average
8.4.2 Couple
Set related parameters according to the coupling relationship.
Auto all: Set Related parameters automatically according to the coupling relationship.
1. Freq step
Freq step has a coupling relationship with RBW at the zero span, when in none-zero mode, Freq step
have coupling relationship with Span. please refer to the introduction of the "Freq Step".
2. Attenuation
Input attenuation has coupling relationship with Ref Level and the preamp. Please refer to introduction
of "amplitude".
3. RBW
RBW has a coupling relationship with the span. Please refer to the introduction of the "Resolution
Bandwidth".
4. VBW
VBW has a coupling relationship with the span. Please refer to the introduction of "VBW".
5. Sweep time
Sweep time has a coupling relationship with RBW, VBW and span. Please refer to the introduction of
"Sweep Time".
6. Mode Couple
Some of the basic parameters are allowed to be set up in coupled operation between different modes.
If the Center Freq is set to on in one mode, it applies to all other modes that support center freq setting
Loading ...