SSA3000X Plus user manual - Page 123

For SSA3015X-PLUS.

Loading ...
Loading ...
Loading ...
SIGLENT
User Manual 101
6.2.4.2 Sweep
Set the sweep mode in single or continuous, the default is continuous. The corresponding icon of the
sweep will be displayed in the status bar at the left of the screen.
1. Single
Set the sweep mode to “Single”. You can set the sweep times , and execute the set number of scans
every time you press "single time".
2. Numbers
Set the sweeps times for a single sweep. In single sweep mode, the system executes the specified
sweeps times and the number shown on the icon in the status bar at the left of the screen varies with
the process of the sweep.
3. Continue
Set the sweep mode to “Continue”. The character Cont on the parameter icon denotes the analyzer is
sweeping continuously.
If the instrument is in single sweep mode and no measurement function is enabled, press this key
and the system will enter continuous sweep mode and sweep continuously if the trigger conditions
are satisfied.
If the instrument is in single sweep mode and a measurement function is on, press this key and the
system will enter continuous sweep mode and measure continuously if the trigger conditions are
satisfied.
In continuous sweep mode, the system will send a trigger initialization signal automatically and
enter the trigger condition judgment directly after each sweep.
Table 6-13 Sweep Times
Parameter
Explanation
Default
1
Range
1 ~ 99999
Unit
N/A
Knob Step
1
Direction Key Step
1
6.2.4.3 Pause/Resume
Press the pause key to pause after the sweep of current frame is completed; press the continue key to
continue the sweep of the real-time spectrum analyzer if it is in the continuous sweep mode, and if it is
in the single sweep mode, the number of times the spectrum analyzer continues to sweep if the number
of times the single scanning is not completed. Continue without clearing historical data after pause.