VDL (E3.1) 10/2020 Page 98/196
Select the field “Temperature” and enter the desired temperature setpoint.
Confirm entry with Confirm icon.
Select the field “Pressure” and enter the desired pressure setpoint.
Confirm entry with Confirm icon.
When entering a value outside the setting range, the message: “Value outside of limits! (Min:
xxx, Max: xxx)” appears (xxx is a wildcard for the limits of the respective parameter). Press
the Confirm icon and repeat the entry with a correct value.
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press
the Close icon to exit the menu without taking over the entries.
With set-point type “Limit”, adapt the safety controller (chap. 14.2) always when you changed
the temperature set-point. Set the safety controller set-point by approx. 5 °C above the con-
troller temperature set-point.
Recommended setting: Set-point type “Limit with safety controller set-point 5 °C.
As long as the pressure threshold of 100 mbar has not yet been reached, the icon “Pressure
threshold not reached” in the screen header in Normal display. Press the flash icon next to
the information icon to access the corresponding text information “Pressure threshold not
reached” (Information messages, chap. 16.1)
10.2 Direct setpoint entry via Normal display
Alternatively you can also enter the setpoints directly via Normal display.
Normal display. Select the setpoint you want to
change.
Example: “Temperature” entry menu. Enter the
desired setpoint and confirm entry with Confirm
icon.