Binder VDL 23 Vacuum drying

User Manual - Page 159

For VDL 23. Also, The document are for others Binder models: VDL 23-UL, VDL 56, VDL 56-UL, VDL 115, VDL 115-UL

PDF File Manual, 197 pages, Read Online | Download pdf file

VDL 23 photo
Loading ...
Loading ...
Loading ...
VDL (E3.1) 10/2020 Page 159/196
Setting the storage rate or rescaling (minimum and/or maximum) will clear the measured-value memory
and the event list.
NOTICE
Danger of information loss when setting the storage rate or rescaling.
Data loss of measured-value memory and event list.
Change the storage rate or scaling ONLY if the previously registered data is no longer
needed.
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press
the Close icon to exit the menu without taking over the entries.
23. Reference measurements
23.1 Checking the temperature in the inner chamber
The controller display was adjusted in the factory to the temperature in the center of the usable volume
(chap. 28.1). The sensor probe of the reference measuring device was connected tightly to an expansion
rack positioned in the middle of the chamber.
23.1.1 Checking the controller display
Conduct the reference sensor into the inner chamber through the measuring port (12). The port must
be largely vacuum tight to enable a typical operation vacuum for the calibration. Use e.g., a silicone plug
with a hole for the sensor wire.
Fix the sensor to the center of an expansion rack in the middle of the useable volume with adhesive
aluminum tape or thermal conductive paste to ensure sufficient thermal conductivity.
Perform the measurement in a thermally stable condition with 3 expansion racks and empty oven.
Equilibrating time: at least 12 hours.
23.1.2 Checking the spatial temperature exactitude
Fix at least 9 sensor probes on 3 racks with adhesive aluminum tape or thermal conductive paste to
ensure sufficient thermal conductivity
The distance of the sensor probes to all inner chamber walls must be at least 10% of the corresponding
inner chamber dimension (see DIN 12880:2007).
Perform the measurement in a thermally stable condition with 3 expansion racks and empty oven.
Response time: at least 12 hours.
Do NOT use the temperature probe of the reference measurement device without any con-
tact to the expansion rack, i.e., do NOT measure vacuum values!
In case the temperature probe is a thermo element, mount it so it is electrically insulated from
the rack.
If you note an excessive divergence between the controller and reference temperatures, please contact
BINDER Service to calibrate the temperature controller.