Binder VDL 23 Vacuum drying

User Manual - Page 13

For VDL 23. Also, The document are for others Binder models: VDL 23-UL, VDL 56, VDL 56-UL, VDL 115, VDL 115-UL

PDF File Manual, 197 pages, Read Online | Download pdf file

VDL 23 photo
Loading ...
Loading ...
Loading ...
VDL (E3.1) 10/2020 Page 13/196
1.6 Type plate
Position of type plate: left chamber side (seen from front), at the bottom right-hand.
Figure 2: Type plate (example of VDL 115)
Indications of the type plate
(example)
Information
BINDER
Manufacturer: BINDER GmbH
VDL 115
Model designation
Vacuum Drying Oven
Chamber name: Vacuum drying oven
Serial No.
000000000000
Serial No. of the chamber
Built
2020
Year of construction
Nominal temperature
110 °C
230 °F
Nominal temperature
IP protection
20
Type of IP protection acc. to standard EN 60529
Temp. safety device
DIN 12880
Temperature safety device acc. to standard DIN 12880:2007
Class
2.0
Class of temperature safety device
Art. No.
9630-0011
Art. No. of the chamber
Project No.
---
Optional: Special application acc. to project no.
1,60 kW
Nominal power
7,0 A
Nominal current
230 V / 50 Hz
Nominal voltage +/- 10% at the indicated power frequency
230 V / 60 Hz
1 N ~
Current type
Explosion proof inner chamber
Explosion proof inner chamber
Ex classification acc. to 2014/34/EU
Ex II 2/3/- G IIB T3 Gb/Gc/- X
Ex classification according to ATEX Directive 2014/34/EU
Max. temp. of inner chamber surface in
cat. 2: +160 °C
Maximum temperature of the inner chamber surfaces:
160 °C / 320 °F (Device category 2)
Max. temp. of heating chamber
surface in cat. 3: +195 °C
Maximum temperature of the outer surface of the preheating
chamber (outer chamber): 185 °C / 365 °F (Device category
3)
Temp. class T3
Temperature class acc. to IEC 60079-0 for the entire cham-
ber
Symbol on the type plate
Information
CE conformity marking
Explosion protection symbol. Ex classification acc. to ATEX
Directive 2014/34/EU
Nominal temp.
IP protection
Safety device
Class
Art. No.
Project No.
Built
110 °C
230 °F
20
DIN 12880
2.0
9630-0011
2020
1,60 kW / 7,0 A
EXPLOSION PROOF INNER CHAMBER
EX CLASSIF. ACC. TO 2014/34/EU
EX II 2/3/- G IIB T3 GB/GC/- X
MAX.
TEMP. OF INNER CHAMBER
SURFACE IN CAT 2: +160 °C
MAX. TEMP. OF HEATING CHAMBER
SURFACE IN CAT 3: +195 °C
TEMP. CLASS T3
230 V / 50 Hz
230 V / 60 Hz
1 N ~
VACUUM DRYING OVEN
BINDER GmbH
Im Mittleren Ösch 5
78532 Tuttlingen / Germany
www.binder-world.com
VDL 115
E3.1
Serial No. 00000000000000
Made in Germany
Loading ...
Loading ...
Loading ...