VDL (E3.1) 10/2020 Page 179/196
Chamber size
23
56
115
Environment-specific data
VDL noise level (mean value)
dB (A)
40
40
40
Energy consumption at 100 °C / 212 °F
Wh/h
140
180
230
All technical data is specified for unloaded chambers with standard equipment (with aluminum racks) at an
ambient temperature of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/- 10
%. Technical data is determined in accordance to BINDER Factory Standard Part 2:2015 and DIN
12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
28.4 Equipment and options (extract)
To operate the VDL vacuum drying oven, use only original BINDER accessories or accesso-
ries / components from third-party suppliers authorized by BINDER. The user is responsible
for any risk arising from using unauthorized accessories.
Regular equipment
Microprocessor display program controller MB2 with touch screen
Ethernet interface for computer communication
USB interface
Safety controller (safety device class 2 according to DIN 12880:2007)
Universal connection for inert gas / ambient air GAS/AIR, adapter with 8 mm hose olive, with solenoid
valve for inert gas / ambient air
Pressure regulator (rear) for sweeping with compressed air
Analog pressure display (manometer) for sweeping with compressed air
Pressure switch for heating release at 100 mbar
Measuring connection (DN 16), rear
2 vacuum expansion racks
Safety glass panel
Options / accessories
Expansion racks , aluminum or stainless steel 1.4571
Lockable door
Analog outputs 4-20 mA for temperature and pressure
Additional universal connection for inert gas / ambient air „GAS/AIR 2“, adapter with 8 mm hose olive,
with solenoid valve for inert gas / ambient air
Additional universal access port 40 mm
FKM door gasket (temperature resistant up to 150 °C / 302 °F
Object temperature display with flexible Pt 100 temperature sensor
APT-COM™ 4 Multi Management Software for logging and display of temperature data and networking
up to 100 chambers with PC
ATEX connection kit for VP4 vacuum pump with various small flange parts
Pump module
Chemical membrane pump VP4 with separator and emission condenser
Ready to connect: Suction power
m
3
/h
1,9
End vacuum
mbar
12
Electrical connection (50-60 Hz)
V
120 / 230
Loading ...
Loading ...
Loading ...