Binder VDL 23 Vacuum drying

User Manual - Page 92

For VDL 23. Also, The document are for others Binder models: VDL 23-UL, VDL 56, VDL 56-UL, VDL 115, VDL 115-UL

PDF File Manual, 197 pages, Read Online | Download pdf file

VDL 23 photo
Loading ...
Loading ...
Loading ...
VDL (E3.1) 10/2020 Page 92/196
DANGER
Explosion or implosion hazard and danger of poisoning by introduction of unsuita-
ble loading material.
Serious injury or death from burns and / or explosion pressure or by poisoning.
Do NOT introduce any substance with an auto-ignition temperature below 200 °C /
392 °F into the chamber.
Do NOT introduce any substance which is combustible or explosive at working temper-
ature into the chamber.
Do NOT introduce energy sources such as batteries or lithium-ion batteries into the
chamber.
Do NOT introduce any substance, which fall under the explosives law into the chamber
Do NOT introduce any combustible dust into the chamber.
Do NOT introduce any substance which could lead to release of toxic gases into the
chamber.
If the auto-ignition temperature of a solvent contained in the drying material is exceeded during the drying
process, there is an immediate risk of fire and explosion. Make sure that the auto-ignition temperature of a
solvent contained will NEVER be reached.
DANGER
Fire and explosion hazard when exceeding the auto-ignition temperature of the sol-
vent.
Serious injury or death from burns and / or explosion pressure.
Take the auto-ignition temperature from the safety specifications of the solvent. In the
case of solvent mixtures, use the lowest auto-ignition temperature.
Correctly set the maximum drying temperature (safety temperature) on the temperature
safety device according to the information panel “Temperature setting”.
Introduce measures to prevent mixing up the material to be dried.
Before starting a new drying process with a different auto-ignition temperature, check
the auto-ignition temperature with the safety specifications of the solvent.
If, during loading material containing solvents falls down, it is possible that the solvent or solvent vapors
will spread outside the loading area, e.g. to the side of or below the chamber. The chamber must be dis-
connected from the power immediately (pull out the power plug or operate e.g., a customer's emergency
stop switch).
The chamber is not to be installed and operated in potentially explosive areas.
DANGER
Explosion hazard due to explosive mixtures in the vicinity of the equipment.
Serious injury or death from burns and / or explosion pressure.
Ensure that NO solvents or explosive solvent-air mixtures are located in the area
around the chamber.
Disconnect the chamber from the power immediately if there is explosive atmosphere in
the area around the chamber outside the indicated loading area.