Binder VDL 23 Vacuum drying

User Manual - Page 40

For VDL 23. Also, The document are for others Binder models: VDL 23-UL, VDL 56, VDL 56-UL, VDL 115, VDL 115-UL

PDF File Manual, 197 pages, Read Online | Download pdf file

VDL 23 photo
Loading ...
Loading ...
Loading ...
VDL (E3.1) 10/2020 Page 40/196
The grounding plan / equipotential bonding prevents sparking
The operator is responsible for the correct installation. He must ensure active extraction (technical
ventilation) when emptying the condensate catchpot of the pump (with or without a pump module)
Spring-mounted safety glass panel
The VDL vacuum drying oven is equipped with a large-surface area safety valve. The inspection window,
consisting of ESG glass (toughened safety glass), is adjustable spring-mounted and serves as a safety
valve in the event of explosion. The laminated safety glass pane provides splinter-protection.
Further measures to prevent accidents
Indications on the type plate
For Ex classification, temperature classes and electrical data, please refer to the operating manual
Chap. 1.6.
Operating manual
An operating manual is available for each chamber.
Temperature monitoring; safety, measurement, and control equipment
The chamber is equipped with a temperature display, which can be read from outside.
The safety, measuring, and control equipment is easily accessible.
Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electrical
equipment (e.g. electric motors, power cables, solenoids). The machine has no permanent magnets. If
persons with active implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of field
source to implant) of 30 cm, an influence of these implants can be excluded with high probability.
Safety controller (temperature safety device class 2)
Temperature limitation via the controller by the safety controller serves to protect the vacuum drying
oven, its environment and loading material against impermissible temperature exceeding., that could
pose a fire risk. If the entered safety controller temperature is exceeded. The heater is turned off at all
poles until manual reset (class 2).
The safety controller must be checked according to its function at appropriate intervals. Test: Set the
safety controller value below the temperature set-point. The chamber must not reach the temperature
set-point, but must turn off the heater when the safety controller value is reached.
Recommended setting: safety controller modeLimit”.
Visual and additionally activatable audible (buzzer) signals indicate temperature exceeding.
Protection against touchable surfaces
Tested according to EN ISO 13732-1:2008.
Floors
See operating manual chap. 5 for correct installation.
Cleaning
See operating manual chap. 25.
Maintenance
For maintenance instructions for the user, please refer to operating manual chap. 26.
Detailed instructions are included in the service manual for this chamber.