VDL (E3.1) 10/2020 Page 164/196
Prior to renewed startup, allow the chamber to completely dry after all cleaning and decontamination
measures.
DANGER
Electrical hazard by water entering the chamber.
Deadly electric shock.
Make sure that the chamber will not become wet during cleaning or decontamination.
Do NOT spill water or cleaning agents over the inner and outer chamber surfaces.
Do NOT put ANY cleaning aids (cloth or brush) into slots or openings on the chamber.
Make sure that no water will enter into slots or openings on the chamber, especially in
the area of the door hinges.
Before cleaning, disconnect the power plug. Let the chamber cool down to ambient
temperature.
Completely dry the chamber before turning it on again.
Avoid electrostatic charges.
DANGER
Explosion hazard due to electrostatic charge when rubbing with a dry cloth.
Serious injury or death from burns and / or explosion pressure.
Use only damp cloth when wiping the device. You may use water and the specified
cleaning agents.
NEVER wipe the device with a dry cloth.
25.2 Cleaning
Disconnect the oven from the power supply before cleaning. Disconnect the power plug.
The interior of the chamber must be kept clean. Thoroughly remove any residues of the
charging material
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces,
inner chamber,
door gaskets
Standard commercial cleaning detergents free from acid or halides.
Alcohol based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Expansion racks,
rack holders
Standard commercial cleaning detergents free from acid or halides, no salt so-
lution or chlorinated solvents.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Controller housing (tri-
angular instrument box)
Standard commercial cleaning detergents free from acid or halides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Zinc coated hinge parts
rear chamber wall
Standard commercial cleaning detergents free from acid or halides.
Do NOT use a neutral cleaning agent on zinc coated surfaces.
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or the
charging material. If there is doubt regarding the suitability of cleaning products, please contact BINDER
service.