Section 2 - Installation and Operation
84
Figure 168. Transmitter Software, 4-20 Monitor and Option Version
TransmitterDataisalsousedtoupdatethecongurationofthe
XNX when an option board is added or changed. To add the new
option, use the
switches to navigate to the Option Revision
screen, then swipe the
magnetic switch on the front panel to
display the Accept New Option Screen. The screen will show
the current option (if any) and the newly installed option. Use the
switches to highlight the option then swipe the
magnetic
switch on the front panel to accept the change. The transmitter
will update the part number of the unit. The new option will then
be operational.
Figure 169. Updating The XNX for Option Boards Added or Changed
?
Transmitter Status
Figure 170. Transmitter Status Screen
Transmitter Status displays information about the XNX unit
including temperature, 4-20 mA output value, and supply
voltage.