XNX Universal Transmitter Technical Manual - Page 133

For XNX UNIVERSAL.

Loading ...
Loading ...
Loading ...
XNX Universal Transmitter
Section 6 - Specications
131
6.2.2 EC Sensor Performance Data, Factory Mutual Veried (see Section 6.3)
Gas Cartridge P/N
Selectable Full
Scale Range (Dis-
play and 4-20mA
Full Scale)
Default
Range
Range
Incre-
ments
Lower
Alarm
Limit
Lower
Detection
Limit
Lower
Explosive
Limit
(% Vol)
Zero
Deviation
Selectable
Cal Gas
Range
Default
Cal Point
Response
Time (T50)
sec
Response
Time (T90)
sec
Accuracy
Operating Temperature Operating Humidity
Min Max Min Max
O
2
Oxygen XNXXSO1FM n/a 23.0% Vol n/a 5.0%Vol 5% Vol n/a n/a
20.9 %Vol
(fixed)
20.9 %Vol T20 <10 <30 <+/-0.5 %Vol -30°C / -34°F 55°C /131°F 15% RH 90% RH
H
2
S Hydrogen Sulfide XNXXSH1FM 10.0 to 50.0 ppm 15.0 ppm 0.1 ppm 5.0 ppm 1.5 ppm n/a -2.5 ppm
30 to 70%
of the
selected full
scale range
10 ppm <20 <30
2 ppm or 10% of
reading, whichever is
greater
-40°C / -40°F 55°C / 131°F 15% RH 90% RH
CO Carbon Monoxide XNXXSC1FM
1
100 to 1000 ppm 300 ppm 100 ppm 30 ppm 15 ppm na/ -25 ppm 100 ppm <15 <30 See footnote 1 -40°C / -40°F 55°C / 131°F 15% RH 90% RH
FOOTNOTES:
1. XNXXSC1FM accuracy over temperature <±10% of reading 20°C/68°F to 55°C/131°F, <±20% of reading 20°C/68°F to -10°C/14°F, <±30% of reading -10°C/14F to -20°C/-4°F.
Recalibration is recommended if the temperature of the local environment has varied by more than -30°C.
NOTES:
• Performanceguresaremeasuredbytestunitscalibratedat50%offullscale,atambientconditionsof20°C,50%RH,withtheECweatherproofcoverattached
• IP rating of FM Cartridges is IP63.
• Barometric pressure effects on the O
2
sensor: The output from the O
2
sensor has pressure effects of <0.1% change of output per % change in pressure. When the barometric
pressure changes by ±20% the output from the O
2
sensorchanges<±0.4%Vol.However,theoxygensensorshowstransientbehaviorwhensubjectedtoarapidchangein
ambientpressureduetoeitherweatheroraltitude.Forexample,a10KPainstantaneouspositivepressurestepchangemaycauseanoverscalealarmconditionforaperiod
of about 12 seconds.
• OperatingtheXNXECsensoratextendedtemperaturerangesforaprolongedtimeperiodexceeding12hoursmycausedeteriorationinthesensorperformanceandshorten
sensorlife.ExtendedtemperaturerangeforXNXECsensorsis-40°Cot-20°C.
• Responsetimesmayincreaseatlowertemperatures.
• FM performance verification is limited to the requirements of the standards identified in Table 6.3 for each cartridge.
• ContactHoneywellAnalyticsforadditionaldataordetails.