XNX Universal Transmitter
Section 5 - Warnings/Faults
112
NOTES
Note 3:
Subtypes Decimal Description
Fault 2 Event
Bits
1
CRC error in safety critical RAM block
2
Error reloading safety critical RAM block from EEPROM
4
Error loading data from Personality board
8
Excel signal level has been below the low signal level
threshold for at least 24 hours
16
Excel beam blocked
32
Personality board error code > 0
64
Option board error code > 0
128
IR mA input > 1 mA and < 3.4 mA
256
IR mA input < 1.0 mA
512
IR forced 10 mA not within +/-1 mA
1024
gains from PGA don't match local copy
2048
error reading or writing EEPROM
4096
ECC reex failure
8192
RAM test failure
16384
Program memory CRC failure
32768
Op code test failure
Fault 3 Event
Bits
1 Interrupt integrity test failure
Note 4:
Optima and Excel fault and warning codes are displayed in the Event History data eld.