9102S Dry-Well Calibrator
Interface Commands
32
7.2 Interface Commands
The various commands for accessing the calibrator functions via the digital interface
are listed in this section (see Table 4 on this page). These commands are used with the
RS-232 serial interface. The commands are terminated with a carriage-return character
(CR, ASCII 13). The interface makes no distinction between upper and lower case
letters, hence either may be used. Commands may be abbreviated to the minimum
number of letters which determines a unique command. A command may be used to
either set a parameter or display a parameter depending on whether or not a value is
sent with the command following a “=” character. For example “s” returns the current
set-point and “s=120.0” sets the set-point to 120.0 degrees.
In the following list of commands, characters or data within brackets, “[” and “]”, are
optional for the command. A slash, “/”, denotes alternate characters or data. Numeric
data, denoted by “n”, may be entered in decimal or exponential notation. Characters
are shown in lower case although upper case may be used. Spaces may be added
within command strings and will simply be ignored. Backspace (BS, ASCII 8) may be
used to erase the previous character. A terminating carriage return (CR, ASCII 13) is
implied with all commands.
Table 4 Controller communications commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point s[etpoint] s set: 999.99 {C
or F}
set: 75.00 C
Set current set-point to n s[etpoint]=n
t[emperature]=n
s=100
t=100
–10 to 122°C
14 to 252°F
Read temperature t[emperature] t t: 999.9 {C or F} t: 55.6 C
Read temperature units u[nits] u u: x u: C
Set temperature units: u[nits]=c/f C or F
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to
Fahrenheit
u[nits]=f u=f
Read scan mode sc[an] sc sc: {ON or OFF} sc: ON
Set scan mode sc[an]=on/off sc=on ON or OFF
Read scan rate sr[ate] sr srat: 99.9 {C or
F}/min
srat:12.4 C/min
Set scan rate sr[ate]=n sr=1.1 0.1 to 99.9°C
0.2 to 179.8°F
Secondary Menu
Read proportional band setting pr[op-band] pr pb: 999.99999 pb: 15.9
Set proportional band to n pr[op-band]=n pr=8.83 0.1 to 30°C
0.2 to 54°F
Read heater power (duty cycle) po[wer] po po: 999.9 po: 6.5
Conguration Menu
Operating Parameters Menu
Read High Limit hl[imit] hl hl: 9999 hl: 125
Set High Limit hl[imit]=n hl=100 50 to 125°C
122 to 257°F