29
Controller Operation
Calibration Parameters
To change the ALPHA setting, press the up or down arrows.
38600 NewALPHAsetting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewALPHAsetting
6.12.4 DELTA
This probe parameter characterizes the curvature of the resistance-temperature
relationship of the sensor. The value of this parameter is set at the factory for best
instrument accuracy.
DeLta FlashesforonesecondandthentheDELTAsettingis
displayed
0.0000 CurrentDELTAsetting
To change the DELTA setting, press the up or down arrows.
0.1000 NewDELTAsetting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewDELTAsetting