Quick Start
AC Power Operation
Quick Start3
Unpacking3.1
Unpack the dry-well carefully and inspect it for any damage that may have occurred
during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
9102S Dry-well
Power Cord
Users Guide with Report of Calibration
RS-232 Cable
9930 Interface-it Software
3102-3 Insert, 3/16”
3102-4 Insert, 1/4”
Insert Removal Tool
Set-up3.2
Place the calibrator on a at surface with at least 6 inches of free space around the
instrument. Always leave enough clearance in front of the instrument to allow for
safe and easy insertion and removal of probes. The prop stand may be swung down to
raise the front of the instrument from a horizontal position. Plug the power cord into a
grounded mains outlet. Observe that the nominal voltage corresponds to that indicated
on the calibrator.
Turn on the power to the calibrator by toggling the power switch on. The fan should
begin quietly blowing air through the instrument and the controller display should
illuminate after 3 seconds. After a brief self-test the controller should begin normal
operation. If the unit fails to operate, check the power connection.
The display should show the well temperature and the well heater will bring the tem-
perature of the well to the set-point temperature.
After using the calibrator, allow the well to cool by setting the temperature to 25°C
and waiting 1/2 hour before turning the instrument off.
AC Power Operation3.3
Plug the dry-well power cord into a mains outlet of the proper voltage, frequency, and
current capability. Refer to Section 3.1, Specications, for the power details. Turn the
dry-well on using the switch on the rear panel. The dry-well will turn on and begin to
heat to the previously programmed temperature set-point. The front panel LED display
will indicate the actual dry-well temperature.