25
Controller Operation
Serial Interface Parameters
Controller Conguration6.8
The controller has a number of conguration and operating options and calibration
parameters which are programmable via the front panel. These are accessed from the
secondary menu after the proportional band function by pressing “SET”. Pressing
“SET” again enters the rst of three sets of conguration parameters — operating
parameters, serial interface parameters, and calibration parameters. The menus are
selected using the up and down arrows and then pressing “SET”.
6.9 Operating Parameters
The operating parameters menu is indicated by,
PAr Operating parameters menu
The operating parameters menu contains the High Limit parameter.
High Limit6.10
The High Limit parameter adjusts the upper set-point temperature. The factory default
and maximum are set to 125°C (257°F). The minimum setting is 50°C (122°F). For
safety, a user can adjust the High Limit down so the maximum temperature set-point is
restricted.
HL FlashesHLandthendisplaysthesetting
125 Flashesthecurrentvalueandthendisplaysthevaluefor
adjustment
125 CurrentHighLimitsetting
Press the u or d to adjust the setting.
100 NewHighLimitsetting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewHighLimitsetting
Serial Interface Parameters6.11
The serial RS-232 interface parameters menu is indicated by,
SErL SerialRS-232interfaceparametersmenu
Press “SET” to enter the menu. The serial interface parameters menu contains param-
eters which determine the operation of the serial interface. The parameters in the menu
are — baud rate, sample period, duplex mode, and linefeed.