9102S Dry-Well Calibrator
Authorized Service Centers
6
correcting a malfunction. You must have a copy of the most recent Report of
Calibration to restore the calibration parameters.
DO NOT
operate this instrument in an excessively wet, oily, dusty, or dirty
environment. Always keep the well and inserts clean and clear of foreign
material.
The dry-well is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with care.
Always carry the instrument in an upright position to prevent the probe sleeves
from falling out. The convenient handle strap allows for one hand carrying.
If a mains supply power uctuation occurs, immediately turn off the instrument.
Power bumps from brown-outs could damage the instrument. Wait until the
power has stabilized before re-energizing the instrument.
The prop stand was not designed to be used as a handle for carrying the
instrument. To avoid damage, do not force the prop stand beyond the incline
positions of the instrument.
Allow for probe expansion inside the well as the block heats.
Most probes have handle temperature limits. Be sure that the probe handle
temperature limit is not exceeded in the air above the instrument.
1.4 Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate service
on your Fluke product: