17
General Operation
Changing Display Units
General Operation5
Setting the Temperature5.1
Section 6.2, Temperature Set-point, on page 19 explains in detail how to set the
temperature set-point on the calibrator using the front panel keys. The procedure is
summarized here.
Press “SET” twice to access the set-point value.1.
Press 2. u or d to change the set-point value.
Press “SET” to program in the new set-point.3.
Press and hold “EXIT” to return to the temperature display.4.
When the set-point temperature is changed the controller switches the well heater on
or off to raise or lower the temperature. The displayed well temperature gradually
changes until it reaches the set-point temperature. The well may require 5 to 10 min-
utes to reach the set-point depending on the span. Another 5 to 10 minutes is required
to stabilize within ±0.05°C of the set-point. Ultimate stability may take 15 to 20 min-
utes more of stabilization time.
Changing Display Units5.2
This instrument can display temperature in Celsius or Fahrenheit. The temperature
units are shipped from the factory set to Celsius. There are two ways to change to
Fahrenheit or back to Celsius.
1 - Press the “SET” and u simultaneously. The temperature display changes units.
or
1 - Press the “SET” key three times from the temperature display to show
Un= C
2 - Press the u or d key to change units.
3 - Press “SET” to save the setting or “EXIT” to continue without changing the
setting.