Specifications
57
DC mA Measurement and Source
Range Resolution
Accuracy,
(% of Reading + Counts)
24 mA 0.001 mA
0.02 % + 2
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ±0.005 % of range per °C
Drive capability: 1000 at 20 mA
Ohms Measurement
Accuracy ± *
Ohms Range
4-Wire 2- and 3-Wire
0 to 400 0.1 0.15
400 to 1.5 k 0.5 1.0
1.5 to 3.2 k 11.5
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ±0.005 % of range per °C
Excitation Current: 0.2 mA
Maximum input voltage: 30 V
* 2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance
not exceeding 100 .
Ohms Source
Ohms Range
Excitation Current from
Measurement Device
Accuracy
±
15 to 400 0.15 to 0.5 mA 0.15
15 to 400 0.5 to 2 mA 0.1
400 to 1.5 k 0.05 to 0.8 mA 0.5
1.5 to 3.2 k 0.05 to 0.4 mA 1
Temperature coefficient -10 °C to 18 °C, +28 °C to
55 °C: ± 0.005 % of resistance range per °C
Resolution
15 to 400 0.1
400 to 3.2 k 1
Frequency Measurement
Range Resolution Accuracy
2.0 to 1000.0 CPM 0.1 CPM ± (0.05 % + 1 count)
1 to 1000 Hz 0.1 Hz ± (0.05 % + 1 count)
1.0 to 10.0 kHz 0.01 kHz ± (0.05 % + 1 count)
Sensitivity: 1V peak-to-peak minimum
Waveform: squarewave
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...