9
Specications and Environmental Conditions
Environmental Conditions
2 Specications and Environmental Conditions
2.1 Specications
Range 35°C to 375°C (95°F to 707°F)
Accuracy ±0.25°C at 50°C
±0.25°C at 100°C
±0.5°C at 375°C
Stability ±0.07°C at 50°C
±0.1°C at 100°C
±0.3°C at 375°C
Resolution 0.1°C or °F
Well-to-Well Uniformity ±0.2°C with sensors of similar size at equal depths within wells
Heating Times 35 to 375°C: 9.5 minutes
Stabilization 5 minutes
Cooling Times 375 to 100°C: 14 minutes
Well Depth 4 inches (102 mm)
1/16 inch (1.6 mm) hole is 3.5 inches (89 mm) deep
Fixed Block Options See Section , Constant Temperature Block Assembly, on page and Figure
on page
Power 115 VAC (±10%), 55–65 Hz, 1.5 A or 230 VAC (±10%), 0.8 A, 45–55 Hz, 175
W
Size 2.25” H x 4.9” W x 6.1”D
(57mm x 125mm x 150 mm)
Weight 2 lb. 3 oz. (1.08 Kg)
Safety Conforms to EN61010-1
Conforms to CAN/CSA C22.2 No.1010.1
UL3111 and ANSI/ISA-S82.01
OVERVOLTAGE (Installation) CATEGORY II, Ploution Degree 2 per
IEC1010-1
Fault Protection Sensor burnout protection, over-temperature cutout, and electrical fuses
Fuse Rating 250 V 3 A FF (very fast acting) NO USER SERVICEABLE PARTS
2.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trouble-free
operation, it must be handled with care. The instrument should not be operated in an
excessively dusty or dirty environment. Maintenance and cleaning recommendations
can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5–50°C (41–122°F)
ambient relative humidity: 15–50%
pressure: 75kPa–106kPa
mains voltage within ±10% of nominal
vibrations in the calibration environment should be minimized
altitudes less than 2000 meters
indoor use only