Parts and Controls
Rear Panel
Parts and Controls4
The user should become familiar with the dry-well calibrator and its parts: (See Figure
2 on this page and Figure 3 on next page).
Rear Panel4.1
Power Cord - The removable power cord, (Figure ) attaches to the back side of the
instrument. It plugs into a standard 115 VAC (optional 230 VAC) grounded socket.
Figure 2 Back Panel
DC Power Jack - The calibrator can be used with a DC power source. The DC input
jack requires 12V and 3.3 amps. See Figure and the inset of Figure for pinout.
Power Switch - The power switch is located on the back panel of the instrument. The
switch is either on or off. The on position is for normal operation. The off position
disconnects power to the entire unit.
Fan - The instrument utilizes a variable speed fan. Under certain circumstances, the
fan may turn off. The fan shuts off at 100°C and above. Slots at the top and around the