FLUKE 9102S/725-KIT Lo-Temp Dry-Well and Process Calibrator Kit

User Manual for the Fluke 9102S/725-KIT Lo-Temp Dry-Well and Process Calibrator Kit - Page 150

For 9102S/725-KIT.

PDF File Manual, 165 pages, Read Online | Download pdf file

9102S/725-KIT photo
Loading ...
Loading ...
Loading ...
9100S Dry-Well Calibrator
Controller Conguration
24
to lower the temperature back down. In this way the temperature is maintained at a
fairly constant level.
The temperature stability of the well and response time depend on the width of the
proportional band. If the band is too wide, the well temperature deviates excessively
from the set-point due to varying external conditions. This is because the power
output changes very little with temperature and the controller cannot respond very
well to changing conditions or noise in the system. If the proportional band is too
narrow, the temperature may swing back and forth because the controller overreacts to
temperature variations. For best control stability, the proportional band must be set for
the optimum width.
The proportional band width is set at the factory and printed on the Report of
Calibration. The proportional band width may be altered by the user if desired to
optimize the control characteristics for a particular application.
The proportional band width is easily adjusted from the front panel. The width may
be set to discrete values in degrees C or F depending on the selected units. The
proportional band adjustment is accessed within the secondary menu. Press “SET” and
“EXIT” to enter the secondary menu and show the heater power. Then press “SET” to
access the proportional band.
S+E
Accessheaterpowerinsecondarymenu
SEC FlashesSECforsecondarymenuandthendisplaysthe
heaterpower
12.0P Heaterpowerinpercent
S
Accessproportionalband
PrOP Flashespropandthendisplaysthesetting
4.1 Proportionalbandsetting
To change the proportional band press u or d.
10.0 Newproportionalbandsetting
To accept the new setting press “SET”. Press “EXIT” to continue without storing the
new value.
S
Acceptthenewproportionalbandsetting
6.8 Controller Conguration
The controller has a number of conguration and operating options and calibration
parameters which are programmable via the front panel. These are accessed from the
secondary menu after the proportional band function by pressing “SET”. Pressing
“SET” again enters the rst of three sets of conguration parameters — operating
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...