Controller Operation
Calibration Parameters
FULL Currentduplexmodesetting
The mode may be changed using the up or down arrows (u d).
HALF Newduplexmodesetting
Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the
next parameter in the menu.
Linefeed6.11.4
The nal parameter in the serial interface menu is the linefeed mode. This parameter
enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after
transmission of any carriage-return. The linefeed parameter is indicated by,
LF Flashesforonesecondandthentheseriallinefeedsetting
is displayed
On Currentlinefeedsetting
The mode may be changed using the up or down arrows (u d).
OFF Newlinefeedsetting
Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the
next parameter in the menu.
6.12 Calibration Parameters
The operator of the instrument controller has access to a number of the calibration
constants: R0, ALPHA, and DELTA. These values are set at the factory and must
not be altered. The correct values are important to the accuracy and proper and safe
operation of the instrument. Access to these parameters is available to the user so that
in the event that the controller memory fails the user may restore these values to the
factory settings. The user should have a list of these constants and their settings with
the instrument manual.
CAUTION: DO NOT change the values of the instrument calibration constants
from the factory set values. The correct setting of these parameters is important
to the safety, proper operation, and performance of the instrument.
The calibration parameters menu is indicated by,
CAL Calibrationparametersmenu
Press “SET” ve times to enter the menu. The calibration parameters menu con-
tains the parameters, Hard Cutout, R0, ALPHA, and DELTA, which characterize the
resistance-temperature relationship of the platinum control sensor. These parameters
may be adjusted to improve the accuracy of the calibrator.